In Stage 1, Set to 2 min at 25°C; 1 Rep.
In Stage 2, Set to 15 min at 55°C; 1 Rep.
In Stage 3, Set to 2 min at 95°C, 1 Rep.
In Stage 4, Step 1 set to 10 sec at 95°C.
In Stage 4, Step 2 set to 30 sec at 60°C.
In Stage 4, Reps should be set to 45 cycles.
In Stage 5, Hold at 25°C forever.